4. Carefully lower the Portable into place, ensuring that the fill connectors are
properly aligned.
5. Place one or both hands on top of the Portable and press straight down
directly over the fill connector , being careful not to depress the release
button on the Reservoir when engaging (Figure 2). This will lower the
Portable unit approximately 3/8 inch (10 mm). Continue to apply downward
force to ensure the fill connectors remain engaged.
6. While holding the Portable unit in the fill position, pull down and hold the
vent valve lever in the open position (90° from the normal “off” position)
(Figure 2). A hissing noise should be noticeable. During filling, maintain a
slight downward force on the unit with one hand to keep the unit steady and
maintain proper filling position.
NOTE: Fill time is usually between 25 and 60 seconds when the Portable is
empty. The amount of time to fill the unit depends on the amount of oxygen
remaining in the Portable.
WARNING
Do not leave the HELiOS Portable unattended during the
filling operation.