26 701641 Rev. E
5. The Portable vent valve does not close properly at the end of the filling
process.
If the vent valve fails to close and the hissing sound and oxygen vapor cloud
continue, carefully remove the Portable by depressing the release button on
the Reservoir. Venting from the bottom of the Portable will stop in a few
minutes. Allow the unit to warm until you can close the vent valve. The
Portable may require as much as 60 minutes to restore adequate pressure for
accurate oxygen flow. If needed, use an alternate source of oxygen such as a
flow control valve attached to the Reservoir.
6. Frost forms on the oxygen supply line connected to the Reservoir.
Frost on the supply line is most likely the result of a leaky connection or a
split in the line. If the connection at the Reservoir outlet is leaking, keep
tightening the nut until the leak stops. If the supply line is split or damaged,
disconnect the line from the Reservoir and contact your home care provider.
7. The Portable does not disengage easily from the Reservoir after filling.
The Portable and Reservoir fill connectors may have become frozen. DO
NOT USE FORCE. Allow a few minutes for the frozen parts to warm, then
disengage the Portable when the ice has melted. To prevent the units from
freezing together, always wipe the male fill connector on the Reservoir and
the female fill connector on the Portable with a clean, dry cloth before
filling.