12 701641 Rev. E
9. Check the oxygen contents indicator to make sure the Portable unit is filled
to the desired level. The amount of liquid oxygen contained in the Portable is
measured by an internal scale that is built into the unit. To operate the
contents indicator: (1) lift the HELiOS Portable unit by the contents
indicator strap, (2) push the bottom backside of the unit so that it is straight
up and down and (3) observe the green bar that displays the liquid oxygen
contents level inside the clear window (Figure 4).
WARNING
If a minor liquid oxygen leak from the Reservoir fill
connector occurs when you disengage the Portable,
re-engage and disengage the unit to help dislodge any
ice or other obstruction. If the liquid oxygen leak
persists, notify your oxygen supplier.
If a major liquid oxygen leak from the Reservoir fill
connector occurs when you disengage the Portable
unit (that is, a steady stream of liquid oxygen), stay
away from the unit and immediately notify your
oxygen supplier.
NOTE: Immediately after filling, you may hear a slight hiss from the Portable. This
is normal and will stop after a few minutes.
(2)
(1)
(3)
Figure 4