20 701641 Rev. E
Connecting the Portable Unit to the Reservoir Unit
When you no longer wish to leave home or move about for the day, you may
connect the Portable to the Reservoir using the oxygen supply line. This will
reduce wasted oxygen and will improve the efficiency of the Reservoir. To
achieve maximum efficiency from the standard HELiOS Reservoir, you should
connect the Portable to the Reservoir at least 10 hours per 24 hour period. It is
important that the Portable be nearly empty so that the oxygen from the Portable
comes directly from the Reservoir. This will maximize the efficiency of the
system. Connect your Portable to the Reservoir as follows:
1. Thread the oxygen supply line nut onto the Reservoir oxygen supply outlet
as pictured below (Figure 11). Make sure the connection is tight. If a hissing
sound is heard, tighten until the hissing stops.
2. Connect the opposite end of the oxygen supply line to the Portable at the
oxygen supply connector by pushing the supply line into the connector
(Figure 12). The connectors should snap together. If the supply line does not
easily engage into the Portable, push the silver release button on the Portable
connector and try again.